Home >

news ヘルプ

論文・著書情報


タイトル
和文: 
英文:Mechanical stability of a microscope setup working at a few kelvins for single-molecule localization 
著者
和文: 日野原拓也, 濱田裕紀, 中村一平, 松下道雄, 藤芳 暁.  
英文: Takuya Hinohara, yuuki hamada, Ippei Nakamura, Michio Matsushita, SATORU FUJIYOSHI.  
言語 English 
掲載誌/書名
和文: 
英文:Chemical Physics 
巻, 号, ページ Vol. 419        p. 246-249
出版年月 2013年6月20日 
出版者
和文: 
英文: 
会議名称
和文: 
英文: 
開催地
和文: 
英文: 
DOI https://doi.org/10.1016/j.chemphys.2013.02.024
アブストラクト A great advantage of single-molecule fluorescence imaging is the localization precision of molecule beyond the diffraction limit. Although longer signal-acquisition yields higher precision, acquisition time at room temperature is normally limited by photobleaching, thermal diffusion, and so on. At low temperature of a few kelvins, much longer acquisition is possible and will improve precision if the sample and the objective are held stably enough. The present work examined holding stability of the sample and objective at 1.5 K in superfluid helium in the helium bath. The stability was evaluated by localization precision of a point scattering source of a polymer bead. Scattered light was collected by the objective, and imaged by a home-built rigid imaging unit. The standard deviation of the centroid position determined for 800 images taken continuously in 17 min was 0.5 nm in the horizontal and 0.9 nm in the vertical directions.